How to count cells with a hemocytometer

Updated October 30, 2026

Skip the maths: use the Hemocytometer count calculatorOpen calculator →

A hemocytometer is a thick glass slide with a counting grid of known volume. Counting the cells over part of that grid tells you how many cells are in each millilitre of your suspension. With trypan blue, the same count also gives viability.

Step 1: Prepare the sample

Step 2: Load the chamber

Place the coverslip on the hemocytometer, then touch the pipette tip to the edge of the coverslip and let about 10 µL flow under by capillary action. The liquid should fill the chamber without overflowing into the moats.

Step 3: Count

A Neubauer grid has nine large squares of 1 mm × 1 mm. With the 0.1 mm chamber depth, each large square holds 0.1 µL. Count the four corner squares under a 10× objective.

Step 4: Calculate

cells/mL = average cells per large square × dilution factor × 10⁴

viability (%) = live cells ÷ (live + dead cells) × 100

Example: live counts of 52, 48, 45 and 51 in the four corner squares, plus 9 blue cells in total, after a 1:1 trypan blue dilution.

The factor 10⁴ converts the 0.1 µL square volume to 1 mL (1 mL = 1000 µL = 10,000 × 0.1 µL).

Common sources of error

Skip the maths: use the Hemocytometer count calculatorOpen calculator →

Frequently asked questions

Why multiply by 10,000?

Each large square of a standard Neubauer chamber holds 0.1 µL. Multiplying the count per square by 10,000 converts it to cells per millilitre.

What is the dilution factor with trypan blue?

Mixing cells and trypan blue in equal volumes is a 1:1 mix, so the dilution factor is 2. If you also diluted the cells beforehand, multiply the two factors.

What viability is acceptable?

Healthy, actively growing cultures are usually above 90% viable. Many protocols require at least 80–90% before seeding or transfection.